Home >> Product Center >>3130 >> Onelife 31 Capillary Regeneration Kit
Product Detail

Onelife 31 Capillary Regeneration Kit

access
Collect
  • 商品简介
  • 商品说明书

   

Onelife 31 capillary regeneration reagent kit is a cleaning agent specifically designed for cleaning the inner walls of 3130 capillaries (36cm, 50cm). The complete set contains seven different components of chemical cleaning and maintenance solutions. The dirt gradually adhering to the inner wall of the capillary tube due to long-term operation can be completely removed, restoring it to its original state. When in use, there is no need to disassemble the capillary tube, and the cleaning is fully realized by utilizing the functions of the equipment itself, making the operation simple and easy.

Regular use of capillary regeneration reagent kits can effectively extend the service life of capillaries, and proper operation can eliminate the need to replace the accessory, saving equipment operating costs for customers.

The Onelife 31 capillary regeneration reagent kit is completely different from the early manual operation of "acid washing". "Acid washing" can damage the inner wall of the capillary, and usually the capillary is scrapped when washed twice.

The Onelife 31 capillary regeneration kit regularly and effectively maintains the capillaries, ensuring that they do not need to be replaced (except for physical damage).

Operating instructions: Please refer to the Onelife series capillary regeneration reagent kit operating instructions document.

Customer application case:


3130清洗案例.jpg



LOGO带R标.png                                  PRODUCT INFORMATION SHEET



Onelife 31 Capillary Regeneration Kit Operating Manual

Suitable for 3130 gene analyzer


  

1、 Product specifications:

7ml x 4 bottles (including: Buffer 1, Buffer 2,Buffer 3,Buffer 4)

2、 Preface:

The Onelife 31 capillary regeneration reagent kit developed and produced by Wanlaifu Biotechnology (Shenzhen) Co., Ltd. provides customers with a method to fully restore the resolution of faulty capillaries in a fast and simple workflow. It can revitalize the old capillaries on your unit's equipment, prolong their service life, and reduce equipment operating costs.

3、 Recommended use:

This reagent kit can be used for capillaries that begin to lose resolution or have stains in the electrophoresis spectrum. It is recommended to use it after every 1500 runs to keep the capillary tube in optimal operating condition.

4、 Safety precautions:

When handling the buffer provided in this kit, please follow good laboratory practices and wear appropriate protective equipment, including gloves, goggles, and a jacket. If any part of the body comes into contact with the buffer, immediately rinse with plenty of water. If necessary, seek medical assistance.

5、 Preparation materials:

16ml of deionized water, two old POP glue empty bottles and clean thoroughly; Pour 8ml of deionized water into two POP bottles, one each.

6、 3130 operation process:

1. Start the computer and instrument, and start the 3130 software.

2. Empty the existing liquid in the anode buffer cup, water tank, waste tank, and buffer tank placed on the sequencer, and then place the empty tank back in place to accommodate the waste liquid generated during capillary cleaning.

3. Put the pre prepared 8ml bottle of deionized water into the position where the POP gel bottle was originally installed on the sequencer.

4. Select the "Replenish Polymer Wizard" program from the "Wizards" main dropdown menu

On the Step1 interface, select "Different Lot" and click on "Next";

On the Step 2 interface, fill in any number in the Polymer lot number column and any date in the Polymer expiration date column. Click Next;

On the Step3 interface, click "Close Valve" and then click "Next";

On the Step4 interface, click "Prime", wait for 30 seconds, and then click "Next";

Step 5 interface, select "No,", All bubbles are gone”, Click Next;

Step 6 interface, select "No,", All bubbles are gone”, Click Next;

On the Step7 interface, click "Finish" to exit directly.

Note: If an error message prompts during the operation of the Replenish Polymer Wizard program, click on "Clear Errors" and continue.

5. Find "Manual Control" from the left navigation menu and click to enter the manual control interface

Select "Buffer Valve" from the dropdown menu on the left side of Send Defined Command For, select "Close/Open buffer valve" from the Command Name column, select "Close", click "Send Command", and wait for the comments status bar to provide feedback;

Select "Polymer delivery pump" from the dropdown menu on the left side of Send Defined Command For

Select "Home Piston" from the Command Name dropdown, click "Send Command", and wait for the comments status bar to provide feedback.

Select "Move Piston Down" in the Command Name column, enter 37500 in the Value column, click "Send Command", and wait for the comments status bar to provide feedback;

Repeat the operation described in step 5 five more times. ★★★★

When the operation described in step 5 is repeated for the fourth time and completed, the cabin door needs to be opened to observe if there are any bubbles in the glue pump. No bubbles, continue with the fifth repetition and skip to step 6. If there are bubbles, open the cabin door and observe the glue pump after performing the fifth repeated operation:

If the bubbles are still in the glue pump, they have not entered the capillary tube; Perform the next step and start step 6.

The bubbles are not in the glue pump. Please repeat the operation in step 5 again (sixth time) to expel the bubbles that enter the capillary tube and avoid the occurrence of hollow parts caused by bubbles; After completing the sixth repetition, proceed to step 6.

6. Select "Over" from the dropdown menu on the left side of the Send Defined Command For column. Select "Turn On/Off Over" in the Command Name column (some software versions are "Select Over State"), then select "On" in the Value column, click "Send Command", and wait for the comments status bar to provide feedback on OK; Select "Set over temperature" in the Command Name column and set the temperature to "40" ℃. Click "Send Command" and wait for the comments status bar to provide feedback. Attention: If stronger repair and cleaning are required, the temperature can be set at 45 ℃ or 50 ℃.

After waiting for 5 minutes, open the front door of the sequencer, remove the ionized water bottle, and install the Buffer 1 bottle of the capillary repair kit.

8. Repeat steps 4-6 above; After completing step 6, let the instrument stand by and wait for 60 minutes (without taking any action).

9. Open the front door of the sequencer, remove the Buffer 1 bottle and install the Buffer 2 bottle of the capillary repair kit.

10. Repeat steps 4-6 above; After completing step 6, let the instrument stand by and wait for 5 minutes (without taking any action).

11. Open the front door of the sequencer, remove the Buffer 2 bottle and install the Buffer 3 bottle of the capillary repair kit.

12. Repeat steps 4-6 above; After completing step 6, let the instrument stand by and wait for 30 minutes (without taking any action).

13. Open the front door of the sequencer, remove the Buffer 3 bottle and install the Buffer 4 bottle of the capillary repair kit.

14. Repeat steps 4-6 above; After completing step 6, let the instrument stand by and wait for 5 minutes (without taking any action).

15. Open the front door of the sequencer, remove the Buffer 4 bottle, and install a new 8ml bottle of deionized water.

16. Repeat steps 4-6 above; The regeneration and repair operation steps for the 3130 capillary have been completed.

17. Open the front door of the sequencer, take out an 8ml bottle of deionized water and install a new bottle of POP glue. From the "Wizards" main drop-down menu, select "Change Polymer Type Wizard" and follow the normal gluing operation process. Remove the deionized water from the glue pump and capillary to allow the glue to fully enter and fill the capillary. If there are bubbles, follow the exhaust bubble process,

18. Clean the waste liquid in the anode buffer cup, water tank, waste tank, and buffer tank, and then install new water and 1x buffer.

Attention:

The shaking of the reagent in the bottle during transportation may cause bubbles, so it is necessary to refrigerate and let it stand until the bubbles disappear before operating the machine.

Retain a sample board for comparison of experimental data before and after cleaning to verify the repair effect of the regeneration reagent.

The repaired capillary tube is similar to the new capillary tube, requiring 3-5 runs to achieve optimal performance. Comparing the data before and after each board run, it can be found that the randomly occurring peaks will gradually decrease and weaken as the number of runs increases.

The repaired capillary can also perform spatial positioning and spectral correction to obtain the best results.

If a more thorough repair is needed, the entire process can be repeated multiple times.

Due to the removal of residue in the connection area of the capillary joint during the regeneration process, the connection between the capillary and the rubber pump may become loose, requiring manual adjustment and tightening of the capillary joint.

Hazard protection signs and hazard statements:


吞噬图片.png 

H360- Swallowing may cause harm to fertility or the fetus.

H351- Swallowing suspected to cause cancer.

Hazardous Material Safety Control Law: Not included



For Research Use Only. Not for use in diagnostic procedures