Electrophoretic separation gel is called separation gel for short. Onelife OP4 and OP7 series products are separation polymers optimized in many aspects for DNA sequencing and fragment analysis, which can be used for long and short DNA sequencing and fragment analysis.

Independent bubble free packaging ensures reliable and stable results, is simple to use, has lower costs, and saves time.

It has superior reading and parsing capabilities, while providing strong dynamic protection for capillaries, and its performance is superior to similar imported products.


LOGO带R标.png                                  PRODUCT INFORMATION SHEET


Product manual

[Product name]: electrophoresis separation gel.

[Packaging Specification]: Onelife OP7 Separation Adhesive 384 Test/Bag; Onelife OP7 separation adhesive 960 test/bag;

[Intended use]: To separate DNA fragments of known size range during electrophoresis.

【 Main components 】: Composed of polyacrylamide polymer solution.

【 Description 】: This product is a colorless and viscous transparent liquid.

【 Storage conditions and expiration date 】: Store refrigerated in an environment of 2-8 ℃, and the product has an expiration date of 6 months. When the product is removed from the refrigerated storage environment and installed with the instrument, and the ambient temperature of the laboratory where the instrument is located is between 15-25 ℃, the product's shelf life is 14 days; When the ambient temperature of the laboratory where the instrument is located is between 25-35 ℃, the product's shelf life is 7 days; When the ambient temperature in the laboratory where the instrument is located is above 35 ℃, the product's shelf life is 0.5 days.

[Applicable Instrument]: Suitable for ABI 3500Dx and 3500xL Dx gene analyzers in the United States.

The number of runs of different packaging specifications on ABI 3500Dx and 3500xL Dx gene analyzers in the United States:


Packaging bag specifications

Capillary tube electrophoresis channel

Number of runs of gene analyzer

Precautions

384 Test/Bag

8-channel capillary tube

60 injections

When the product is run on the genetic analyzer for more than the number of times it has been used or exceeds the expiration date, if it continues to run, it may affect the quality of detection data or damage the capillary.

24-channel capillary tube

20 injections

960 test/bag

8-channel capillary tube

120 injections

24-channel capillary tube

50 injections

[Interpretation of test results]: When the operating system pops up an error dialog prompt window of "electrophoresis current is too high" during the operation of the gene analyzer, it means that the electrophoresis separation gel installed on the instrument at this time is invalid, please replace it with a new electrophoresis separation gel.

Limitations of the testing method: After electrophoresis separation, the test sample will obtain 700-1000bp of effective read length data. Data above 1000bp can be considered invalid data; When the effective data of a certain channel is below 700bp or the data is invalid, please check the corresponding test sample of that channel before electrophoresis: sample extraction, PCR amplification, purification and other operation steps.

【 Precautions 】:

1. Before use, please check the expiration date on the label to ensure that the product has not expired; And it will not expire during use. Do not use if the product has expired, the packaging bag is damaged, the top sealing strip is lost or damaged.

2. After removing the product from the 2-8 ℃ low-temperature freezer, let it stand in an ambient temperature of 15-30 ℃ and let it naturally heat up to the ambient temperature.

3. Follow the prompts in the software wizard window. The wizard window prompts that after electrophoresis separation and gel loading, tear off the top sealing strip of the package; If a little electrophoresis separation gel appears below the sealing strip after it is opened, it is normal (it is the residual gel produced during filling) and will not cause product performance problems.

4. Slide the plastic head on the top of the electrophoresis separation gel packaging bag into the installation slot of the instrument lever assembly, push the lever assembly handle upward, and snap the plastic head on the top of the bag into the one-way valve of the glue pump assembly.

5. Click Refresh in the software instrument window to check whether the data read by the instrument to the newly installed electrophoresis separation gel is correct.

[Explanation of identification]: 384 or 960 in the product packaging specification means that the packaging size of electrophoresis separation gel can meet the test quantity of 384 or 960 test samples.

[Basic Information]:

Registered person/manufacturer name: Wanlaifu Biotechnology (Shenzhen) Co., Ltd.

Address/Production Enterprise Address: Building A11, Life Science Industrial Park, No. 140 Jinye Avenue, Sanxi Community, Kuiyong Street, Dapeng New District, Shenzhen.

Contact information: (0755) 61347309

Medical Device Production Filing Certificate No.: Yueshen Food and Drug Administration Production Preparation No. 20200150

Medical Device Filing Certificate/Product Technical Requirements Number: Yue Shen Xi Bei 20230269

[Product Validity Period]: See product label

[Approval and revision date of instruction manual]: March 8, 2023

If there are any problems, please contact the production enterprise



For Research Use Only. Not for use in diagnostic procedures